Circular Dichroism for Peptides: Reading Secondary Structure

Circular dichroism (CD) spectroscopy is a quick way to see whether a peptide in solution is helical, sheet-like or largely disordered, and whether that changes with solvent, temperature or time. This guide covers what the spectrum measures, the shapes to recognise, unit conversion, sample preparation, structure estimation and what CD cannot tell you.

General laboratory-science education. Not medical advice. Sealed Peptides materials are for in-vitro and laboratory research use only, and are not for human or veterinary use.

What circular dichroism measures

Circular dichroism is the difference between how much left- and right-circularly polarised light a sample absorbs. Only chiral, asymmetric structures give a signal, and a chain of L-amino acids folded into a regular shape gives a characteristic one. Instruments usually report the difference as an ellipticity in millidegrees (mdeg).

Two wavelength regions are used. In the far ultraviolet, about 190 to 250 nm, the signal comes mainly from the amide bonds of the backbone, so the spectrum reflects secondary structure. In the near ultraviolet, about 250 to 320 nm, aromatic side chains and disulfide bonds contribute, which reports on their surroundings in a folded molecule.[1] The most common uses are to check whether a molecule is folded, whether a change in sequence or conditions alters its conformation, and how stable that conformation is; measurements can be made on samples of 20 µg or less in a few hours.[2]

The spectra to recognise

Each kind of secondary structure has a typical far-UV shape:[2]

  • Alpha-helix. Two negative bands near 222 and 208 nm and a strong positive band near 192 nm. The 222 nm band is the one most often used to follow helix content.
  • Beta-sheet. A single negative band near 216 to 218 nm and a positive band near 195 nm. The exact shape varies with how the strands are arranged and twisted.
  • Disordered and polyproline II. A strong negative band near 195 to 200 nm and little signal above 210 nm, or a weak positive band near 220 nm. Many short linear peptides in water give this spectrum, and much of it has been attributed to the extended polyproline II conformation rather than to a truly random chain.[3]

Real spectra are usually mixtures. A peptide that is partly helical in water gives a spectrum between the helix and disordered shapes, and the 222 nm signal grows as helix content rises.

Units: from mdeg to mean residue ellipticity

Raw ellipticity depends on how much peptide is in the light path, so spectra are converted to mean residue ellipticity, which normalises per amide bond and allows molecules of different lengths to be compared:[1]

[θ] = (θ in mdeg × MRW) / (10 × path length in cm × concentration in mg/mL), in units of deg cm² dmol⁻¹, where the mean residue weight MRW is the molecular weight divided by the number of peptide bonds (residues minus one).

A worked example: a 20-residue peptide of 2200 Da has an MRW of 2200 / 19 ≈ 115.8. At 0.2 mg/mL in a 0.1 cm cell, a reading of -10 mdeg at 222 nm gives (-10 × 115.8) / (10 × 0.1 × 0.2) = -1158 / 0.2 ≈ -5,790 deg cm² dmol⁻¹. The peptide mass calculator gives the molecular weight and residue count.

The concentration enters the result directly, so a 20% error in concentration is a 20% error in [θ] and in any helix estimate built on it. A concentration calculated from the weighed powder includes water and counter-ions (see purity versus net peptide content); a value measured by absorbance with an extinction coefficient or by amino acid analysis is more reliable.[1]

Preparing the sample and buffer

Most problems in CD come from the sample rather than the instrument. Published guidelines stress calibrating the instrument, knowing the concentration accurately and keeping the total absorbance of the sample, cell and solvent low enough that the detector still receives enough light.[1] In practice:

  • Choose a transparent buffer. Low-concentration phosphate works well; chloride, imidazole, high Tris concentrations and many organic additives absorb strongly below about 200 nm. Sodium fluoride can replace sodium chloride when salt is needed. The buffer and pH guide covers choosing the pH itself.
  • Match path length to concentration. Far-UV spectra are typically recorded in cells of 0.1 cm or less at a few tenths of a mg/mL; near-UV spectra need longer cells and higher concentrations.
  • Record a blank. Measure the buffer alone in the same cell and subtract it, and watch the detector voltage so that data where too little light reaches the detector are discarded.
  • Check the sample is clear. Particles and aggregates scatter light and distort the spectrum. Centrifuging or filtering before measurement helps; see the solubility and aggregation guide.

Estimating secondary structure

Deconvolution programs fit a measured spectrum as a combination of reference spectra from molecules of known structure. A comparison of three widely used methods, CONTIN, SELCON3 and CDSSTR, found that they performed comparably despite different algorithms, that the best choice depended on the reference set and wavelength range, and that analysing a spectrum with all three improves reliability.[4] Reference databases and web servers such as DICHROWEB have made these analyses routine, and data extending to lower wavelengths increase the information in a spectrum.[5]

Beta-rich structures are harder. Their spectra vary with the parallel or antiparallel arrangement and the twist of the sheets, and the BeStSel method was built to take this into account; it can distinguish parallel from antiparallel sheets and was applied to amyloid fibrils of disease-related proteins and peptides.[6]

Most reference sets are made of folded proteins, so fractions reported for a short, flexible peptide carry more uncertainty than those for a globular protein. For many peptide questions the shape of the spectrum, the change in the 222 nm signal, or the presence of an isodichroic point in a titration is more informative than a percentage.

Solvents, temperature and aggregation

Many peptides that are mostly disordered in water become helical when a cosolvent such as 2,2,2-trifluoroethanol (TFE) is added. In a study of repeating-sequence peptides of 7 to 22 residues, thermal unfolding curves measured by CD between 0 and 50% TFE fitted a single helix-coil model, suggesting that helix formation is basically the same in TFE-water mixtures as in water; helix propensity rose steadily up to about 25% TFE and then levelled off.[7] A review of TFE and related alcohols concluded that no single mechanism yet accounts for all of their effects on conformation.[8] A TFE spectrum therefore shows what a sequence can adopt, not what it adopts in buffer, and the two should be reported separately.

Following the 222 nm signal while the temperature is raised gives a thermal unfolding curve. The same study found that the ellipticity of a complete helix itself depends strongly on temperature, which has to be considered when interpreting such curves.[7] Spectra recorded over hours or days can also show aggregation: a shift from a disordered spectrum towards a single minimum near 218 nm is a common sign that sheet-rich assemblies are forming.[6]

What CD does not tell you, and what to record

CD reports on the average conformation of everything in the cell. It does not give residue-by-residue information of the kind obtained from crystallography or NMR,[2] and it does not establish identity or purity: an impurity or a degraded form contributes to the spectrum without being visible as a separate signal. Identity and purity are questions for mass spectrometry and HPLC. For a CD result to be interpretable, record:

  • Sample. Sequence, how the concentration was determined, and the buffer, pH and any cosolvent.
  • Measurement. Cell path length, temperature, wavelength range, bandwidth, number of scans and the blank subtracted.
  • Units and analysis. Whether data are in mdeg or mean residue ellipticity, the MRW used, and the deconvolution method and reference set.

This guide describes general laboratory methods and published literature only. It is not a statement about the structure or behaviour of any particular material. Terms used here are defined in the glossary, and questions can be sent to support.

Terms used in this guide

Related tools and guides

References

  1. Kelly SM, Jess TJ, Price NC. How to study proteins by circular dichroism. Biochim Biophys Acta. 2005;1751(2):119-39. PMID 16027053 · DOI
  2. Greenfield NJ. Using circular dichroism spectra to estimate protein secondary structure. Nat Protoc. 2006;1(6):2876-90. PMID 17406547 · DOI
  3. Woody RW. Circular dichroism spectrum of peptides in the poly(Pro)II conformation. J Am Chem Soc. 2009;131(23):8234-45. PMID 19462996 · DOI
  4. Sreerama N, Woody RW. Estimation of protein secondary structure from circular dichroism spectra: comparison of CONTIN, SELCON, and CDSSTR methods with an expanded reference set. Anal Biochem. 2000;287(2):252-60. PMID 11112271 · DOI
  5. Whitmore L, Wallace BA. Protein secondary structure analyses from circular dichroism spectroscopy: methods and reference databases. Biopolymers. 2008;89(5):392-400. PMID 17896349 · DOI
  6. Micsonai A, Wien F, Kernya L, Lee YH, Goto Y, Réfrégiers M, et al. Accurate secondary structure prediction and fold recognition for circular dichroism spectroscopy. Proc Natl Acad Sci U S A. 2015;112(24):E3095-103. PMID 26038575 · DOI
  7. Luo P, Baldwin RL. Mechanism of helix induction by trifluoroethanol: a framework for extrapolating the helix-forming properties of peptides from trifluoroethanol/water mixtures back to water. Biochemistry. 1997;36(27):8413-21. PMID 9204889 · DOI
  8. Buck M. Trifluoroethanol and colleagues: cosolvents come of age. Recent studies with peptides and proteins. Q Rev Biophys. 1998;31(3):297-355. PMID 10384688 · DOI

Cited literature is provided for scientific context only. Many of these reviews concern proteins or pharmaceutical development; they are cited for general chemistry and do not describe or endorse any use of these products.

For laboratory research use only. Not for human or veterinary use. This page is general chemistry education and does not describe or endorse any use of Sealed Peptides products. Cited literature is provided for scientific context only.

Guide FAQ

What does a circular dichroism spectrum show for a peptide?

In the far ultraviolet, roughly 190 to 250 nm, the signal comes mainly from the peptide backbone, so the shape of the spectrum reflects secondary structure. An alpha-helix gives negative bands near 222 and 208 nm and a positive band near 192 nm; a beta-sheet gives a single negative band near 216 to 218 nm; a largely disordered peptide gives a strong negative band near 195 to 200 nm.

How do I convert CD millidegrees to mean residue ellipticity?

Mean residue ellipticity = (ellipticity in mdeg x mean residue weight) / (10 x path length in cm x concentration in mg/mL), where mean residue weight is the molecular weight divided by the number of peptide bonds. For a 20-residue peptide of 2200 Da at 0.2 mg/mL in a 0.1 cm cell reading -10 mdeg, the result is about -5,790 deg cm2 dmol-1.

Which buffers can be used for far-UV CD?

Buffers that absorb little below 200 nm, such as low-concentration sodium phosphate, or sodium fluoride in place of chloride salts, are commonly used. Chloride, imidazole, high concentrations of Tris and many organic additives absorb strongly in the far UV and limit how low in wavelength the spectrum can be read.

Does a helix formed in trifluoroethanol mean the peptide is helical in water?

No. Trifluoroethanol (TFE) stabilises helices in peptides that have some helix-forming tendency, so a TFE spectrum shows what the sequence can adopt rather than what it adopts in water. Studies have used TFE titrations to extrapolate helix-forming properties back to water, but the two conditions should be reported separately.

Who is this guide written for?

Laboratory staff and researchers handling research-use-only materials. "Circular Dichroism for Peptides: Reading Secondary Structure" covers laboratory practice only and describes no use of any product in people or animals.

Does Sealed Peptides give usage advice?

No. Our guides and calculators cover laboratory handling, analysis and unit arithmetic only. All products are sold for laboratory research use only.

Is there a calculator that goes with this guide?

Yes: Peptide mass calculator, Concentration calculator, Peptide properties calculator. They run free in the browser with no signup.

How are lyophilized research peptides usually stored?

Lyophilized powders are generally kept sealed, dry, protected from light and frozen (around -20°C). Each product page lists its stated storage condition, and our peptide storage guide explains the reasoning.

More answers: Site FAQ · Storage guide · Storage reference tables

Explore Sealed Peptides

Coming soon: Epitalon · Thymosin Alpha-1 · Glutathione · ARA-290